X-ray effects on pacemaker type circuits.

作者信息

Blamires N G, Myatt J

出版信息

Pacing Clin Electrophysiol. 1982 Mar;5(2):151-5. doi: 10.1111/j.1540-8159.1982.tb02206.x.

Abstract

Queries have been raised concerning the potential hazards of X-ray irradiation on patients using the new generation of heart pacemakers based on digital circuitry. The present study was undertaken to provide some answers to these queries. The work was conducted in two parts. First, a literature search was done and, second, circuits using current "state of the art" digital technology were irradiated with X-rays. Watch circuits were chosen because of their availability and built-in facilities by which their function could be tested. Doses up to 330 rads were administered to them using energies of 46, 114, and 141 KeV. The conclusion drawn from both parts of the study was that X-rays used for diagnostic purposes were unlikely to affect the performance of this type of circuit in any way. It was accepted that for therapeutic purposes doses far in excess of this are administered and circuit malfunctions are likely to occur. To assess the probability of a digital pacemaker malfunctioning, samples of that particular type would have to be irradiated at the relevant dose.

摘要

对于使用基于数字电路的新一代心脏起搏器的患者而言,X射线辐射的潜在危害已引发诸多疑问。开展本研究旨在为这些疑问提供一些答案。这项工作分两部分进行。首先,进行了文献检索;其次,用X射线对采用当前“先进”数字技术的电路进行了辐照。之所以选择手表电路,是因为其可得性以及具备可用于测试其功能的内置设施。使用46、114和141千电子伏特的能量对其施加高达330拉德的剂量。从研究的两个部分得出的结论是,用于诊断目的的X射线不太可能以任何方式影响此类电路的性能。大家公认,用于治疗目的时所施用的剂量远远超过此值,并且很可能会发生电路故障。为评估数字起搏器发生故障的可能性,必须以相关剂量对该特定类型的样本进行辐照。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验