Suppr超能文献

Endless loop tachycardia in an AV universal [DDD] pacemaker.

作者信息

Furman S, Fisher J D

出版信息

Pacing Clin Electrophysiol. 1982 Jul;5(4):486-9. doi: 10.1111/j.1540-8159.1982.tb02265.x.

Abstract

Implantation of an AV universal [DDD] pacemaker, stimulating and sensing both atrium and ventricle caused a pacemaker mediated tachycardia, begun by a P wave occurring after a premature ventricular contraction, and sustained by P waves caused by retrograde AV conduction. The tachycardia was terminated by application of a magnet to the pulse generator which temporarily made it insensitive to incoming signals, both P and QRS.