Moore R, Korbuly D, Kotula F, Castaneda-Zuniga W, Amplatz K
Rofo. 1982 Sep;137(3):333-40. doi: 10.1055/s-2008-1056212.
The images produced in rotational scanography, currently under investigation at several institutions, may contain several kinds of artifacts. The sources of such artifacts are: 1. vertical and horizontal misalignment of the focal spot with the axis of rotation; 2 misalignment of the primary and secondary slits; 3 a density gradient due to slit cut-off; 4 vibration of the x-ray tube during rotation; 5. A density gradient due to variation in motor speed; 6. radiographic striations due to ripple of the tube output. Striations are present even if a 12-pulse generator is used, but they disappear if a constant potential generator is employed.
目前多家机构正在研究的旋转扫描摄影所产生的图像可能包含多种伪影。此类伪影的来源有:1. 焦点与旋转轴的垂直和水平不对准;2. 主狭缝和副狭缝的不对准;3. 由于狭缝截止导致的密度梯度;4. X射线管在旋转过程中的振动;5. 由于电机速度变化导致的密度梯度;6. 由于管输出的波动导致的射线照相条纹。即使使用12脉冲发生器,条纹也会出现,但如果采用恒压发生器,条纹就会消失。