A quenched-flow apparatus which allows the measurement of the kinetics of a reaction in one stroke.

作者信息

Gangloff J, Pouyet J, Kern D, Dirheimer G

出版信息

J Biochem Biophys Methods. 1984 Jul;9(3):201-13. doi: 10.1016/0165-022x(84)90025-3.

Abstract

A chemical quenched-flow apparatus is described which measures, in a unique stroke, enough data points (8-11) for establishing the kinetics curve of a reaction. Only very small volumes of reaction solutions (2 X 500 microliters) are required. The time intervals between which the kinetic data may be measured range from 5 to 37 ms and from 120 to 450 ms with the corresponding mixing times of 0.6 and 5 ms, respectively. This apparatus was used to investigate the pre-steady-state domain of the aminoacylation reaction of tRNAVal by valyl-tRNA synthetase from yeast.