Constantinou C, Kember N F, Huxtable G, Whitehead C, Stedeford J B, Weatherburn H
Phys Med Biol. 1980 May;25(3):489-99. doi: 10.1088/0031-9155/25/3/008.
The measurement of the physical parameters of a high-energy proton beam, using a range of liquid and solid tissue substitutes, is described. The system, the detectors used and the experimental verification of the tissue equivalence of the new tissue substitutes is presented. The measurements with the scattered but uncollimated proton beam in muscle- and brain-equivalent liquids and in water are compared to similar data obtained from the scattered but collimated beam. The effect of lung, fat and bone on the dose distributions in composite phantoms is also investigated and the necessary corrections established. A simulated patient treatment indicated that the Bragg peak can be positioned with an error not exceeding +/- 0.5 mm.
本文描述了使用一系列液体和固体组织等效物对高能质子束物理参数的测量。介绍了该系统、所使用的探测器以及新组织等效物的组织等效性实验验证。将在肌肉等效液、脑等效液和水中散射但未准直的质子束的测量结果与从散射但准直的束中获得的类似数据进行了比较。还研究了肺、脂肪和骨对复合体模中剂量分布的影响,并确定了必要的校正。模拟患者治疗表明,布拉格峰的定位误差不超过±0.5毫米。