Mattsson L O, Johansson K A, Svensson H
Acta Radiol Oncol. 1982;21(2):139-44. doi: 10.3109/02841868209133997.
A check of the dosimetry using ionization chambers and applying the procedures described in the Nordic dosimetry protocols (NACP 1980, 1981) was carried out using the ferrous sulphate dosimeter as reference. Measurements have been carried out in electron beams in the energy range 2.8 to 27.5 MeV and in 60Co gamma beams. Consistent results were obtained with the two systems within about one per cent. Larger differences were only observed in electron beams containing a large proportion of scattered low energy electrons and contaminating roentgen rays which were not corrected for in the ionization chamber method.
以硫酸亚铁剂量计作为参考,使用电离室并采用北欧剂量测定协议(NACP 1980、1981)中描述的程序对剂量测定进行了检查。在能量范围为2.8至27.5 MeV的电子束和60Coγ射线束中进行了测量。两个系统得到了一致的结果,误差在约1%以内。仅在含有大量散射低能电子和未在电离室方法中校正的污染伦琴射线的电子束中观察到较大差异。