Kerlan R K, Stimac G, Pogany A C, Ring E J
AJR Am J Roentgenol. 1984 Nov;143(5):1085-7. doi: 10.2214/ajr.143.5.1085.
An in vitro model was developed to measure bile flow at varied pressures through several commonly used transhepatic drainage catheters. As expected, larger catheters conduct more bile flow at a given pressure. All catheters studied conduct adequate amounts of bile at normal physiologic pressure gradients. The clinical observation that large catheters (10-12 French) are necessary for adequate antegrade drainage in some patients must be explained by an abnormal bile viscosity, increased duodenal pressures, or reduced biliary secretory pressure.
开发了一种体外模型,以测量通过几种常用的经肝引流导管在不同压力下的胆汁流量。不出所料,在给定压力下,较大的导管传导的胆汁流量更多。所有研究的导管在正常生理压力梯度下都能传导足够量的胆汁。对于某些患者需要大导管(10-12法式)进行充分的顺行引流这一临床观察结果,必须用胆汁粘度异常、十二指肠压力增加或胆汁分泌压力降低来解释。