Suppr超能文献

Particulate contamination in disposable corneal trephines.

作者信息

Pardos G J, Meltzer D W

出版信息

Arch Ophthalmol. 1983 Apr;101(4):562-3. doi: 10.1001/archopht.1983.01040010562005.

Abstract

In two cases scheduled for corneal transplantation, edges of donor corneal buttons were discolored by a black contaminant transferred from the disposable trephines. Trephines from two different manufacturers produced this same problem. Scanning electron microscopy and energy-dispersive x-ray analysis indicated that the contaminant was a mixture of stainless steel particles from the trephine and silicon carbide abrasive used in sharpening.