Accuracy of calibrated-loop transfer.

作者信息

Albers A C, Fletcher R D

出版信息

J Clin Microbiol. 1983 Jul;18(1):40-2. doi: 10.1128/jcm.18.1.40-42.1983.

Abstract

The accuracy of the 0.001-ml calibrated platinum loop was tested by two methods: visible absorption spectroscopy and weight determinations. By both methods it was demonstrated that the volume of the loop delivery is related to the angle at which the loop is withdrawn from the solution being sampled and the diameter of the container, provided that the volume in the container is adequate to cover the loop. Vertical sampling from small-diameter containers (less than or equal to 7-mm inside diameter) delivered approximately 50% of 0.001 ml, and sampling at a 45 degree angle from larger containers (greater than or equal to 22-mm inside diameter) gave a delivery volume of greater than 150% of 0.001 ml, resulting in a threefold difference in the amount delivered and an error rate of +/- 50%.

摘要

通过两种方法测试了0.001毫升校准铂环的准确性:可见吸收光谱法和重量测定法。两种方法均表明,只要容器中的溶液体积足以覆盖铂环,铂环移液的体积就与铂环从采样溶液中抽出的角度以及容器的直径有关。从小直径容器(内径小于或等于7毫米)垂直采样时,移液量约为0.001毫升的50%,而从大直径容器(内径大于或等于22毫米)以45度角采样时,移液量大于0.001毫升的150%,导致移液量相差三倍,误差率为±50%。

相似文献

1
Accuracy of calibrated-loop transfer.校准环路传递的准确性。
J Clin Microbiol. 1983 Jul;18(1):40-2. doi: 10.1128/jcm.18.1.40-42.1983.
6
Effects of phantom volume and shape on the accuracy of MRI BANG gel dosimetry using BANG3.
Br J Radiol. 2008 Jan;81(961):46-50. doi: 10.1259/bjr/71353258. Epub 2007 Nov 12.

引用本文的文献

7
Bronchoscopic diagnosis of pneumonia.肺炎的支气管镜诊断
Clin Microbiol Rev. 1994 Oct;7(4):533-58. doi: 10.1128/CMR.7.4.533.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验