Chakraborty D P, Yester M V, Barnes G T, Lakshminarayanan A V
Radiology. 1984 Jan;150(1):225-9. doi: 10.1148/radiology.150.1.6359264.
The authors tested the image quality and dose savings of self-masking subtraction tomosynthesis (SST), which combines digital tomosynthesis with subtraction of a blurred self-mask. High-quality images of the inner ear of a head phantom were obtained at moderate dose savings. Although they were taken with linear motion, they did not exhibit the streaking due to off-fulcrum objects that is characteristic of conventional linear tomography. SST could reduce patient dose by a factor of at least 12 in examinations of the inner ear, and the mechanical aspects can be implemented with moderate modifications of existing instrumentation.
作者测试了自屏蔽减影断层合成(SST)的图像质量和剂量节省情况,该技术将数字断层合成与模糊自屏蔽的减法相结合。在适度节省剂量的情况下,获得了头部模型内耳的高质量图像。尽管这些图像是通过直线运动拍摄的,但它们并未表现出传统直线断层扫描中因支点外物体而产生的条纹。在内耳检查中,SST可将患者剂量降低至少12倍,并且在对现有仪器进行适度修改的情况下即可实现其机械方面的操作。