Dines K A, Sheets P W, Brink J A, Hanke C W, Condra K A, Clendenon J L, Goss S A, Smith D J, Franklin T D
Ultrason Imaging. 1984 Oct;6(4):408-34. doi: 10.1177/016173468400600403.
Experimental results are presented demonstrating the application of pulse echo ultrasound to imaging the skin. A laboratory prototype B-mode mechanical scanner was employed to obtain images of human skin, both in vitro and in vivo, using broadband pulsed ultrasound at 25 MHz. Images were formed by processing digitized A-mode waveforms and displaying the resulting two-dimensional cross sections using a digital imaging system. Images obtained by rectifying the A-modes are compared to those derived using a software-based cross-correlation technique. Scans of test targets demonstrate that an axial resolution of 100 m can be achieved at 25 MHz when the digital correlation method is employed. Lateral resolution is limited by the 0.25 mm half-power focal beamwidth of the transducer. Seventeen in vitro ultrasonic scans of human skin were compared to frozen section histology. Average skin depth was well correlated between the two techniques ( = 0.99, p less than 0.001). Application of cross-correlated processing to 25 MHz in vivo images produced good delineation of epidermis, papillary, and reticular dermis. Conversion to a 50 MHz transducer did not delineate skin layers as well as the 25 MHz transducer due to inherent difficulties with transducer reverberations.
本文展示了脉冲回波超声在皮肤成像方面的实验结果。使用一台实验室原型B型机械扫描仪,以25兆赫兹的宽带脉冲超声获取人体皮肤的体内和体外图像。通过处理数字化的A模式波形,并使用数字成像系统显示所得的二维横截面来形成图像。将通过对A模式进行整流获得的图像与使用基于软件的互相关技术得到的图像进行比较。对测试目标的扫描表明,采用数字相关方法时,在25兆赫兹频率下可实现100微米的轴向分辨率。横向分辨率受换能器0.25毫米半功率聚焦波束宽度的限制。对人体皮肤进行了17次体外超声扫描,并与冰冻切片组织学进行比较。两种技术测得的平均皮肤深度具有良好的相关性(r = 0.99,p小于0.001)。对25兆赫兹的体内图像应用互相关处理能够清晰描绘表皮、乳头层和网状真皮。由于换能器混响存在固有困难,换成50兆赫兹的换能器后对皮肤层的描绘效果不如25兆赫兹的换能器。