[移动条带照射技术的实验与数学验证]

[Experimental and mathematical check of the moving-strip irradiation technic].

作者信息

Lindner H, Kneschaurek P, Czempiel H, Bogner L, Schwarzer J U, Atzinger A

出版信息

Strahlentherapie. 1983 Jan;159(1):13-7.

PMID:6404014
Abstract

The moving-strip irradiation of the total peritoneal region plays an important role in the adjuvant therapy of the ovarian carcinoma. TLD dosimetric and mathematic investigations of this radiotherapeutic method show that 1. the best dose homogeneity in the longitudinal profile is achieved by 60Co units, if the patients are not too stout, 2. a higher relative depth dose has to be estimated than for an open field of the same total field size, 3. an additional 2,5 cm strip has to be inserted in order to achieve an abrupt drop on the border of the total irradiation field, 4. this method offers a good guarantee against false adjustment, if this is not too grave, 5. the conformity between the dose distribution curves calculated and those measured by TLD is excellent.

摘要

全腹膜区域的移动条带照射在卵巢癌的辅助治疗中起着重要作用。对这种放射治疗方法的热释光剂量测定和数学研究表明:1. 如果患者不过于肥胖,60Co 装置能在纵向轮廓上实现最佳剂量均匀性;2. 与相同总面积的开放野相比,必须估计更高的相对深度剂量;3. 为了在全照射野边界实现剂量陡然下降,必须额外插入一条 2.5 厘米宽的条带;4. 如果误差不太严重,这种方法能很好地保证避免错误调整;5. 计算得到的剂量分布曲线与热释光测量得到的曲线之间的符合度极佳。