Suppr超能文献

CO2 lasers: beam patterns in relation to surgical use.

作者信息

Fava G, Emanuelli H, Cascinelli N, Bandieramonte G, Canestri F, Marchesini R

出版信息

Lasers Surg Med. 1983;2(4):331-41. doi: 10.1002/lsm.1900020406.

Abstract

According to surgeons operating with a variety of CO2 lasers available at the National Cancer Institute of Milan (Coherent, Sharplan, Valfivre), these lasers have different cutting and coagulation properties. To identify what physical parameters might corroborate the subjective impression of the surgeons, a comparative study of the crater forms in perspex samples was performed. Perspex was chosen for its thermal properties (in fact, its thermal conductivity and diffusivity are similar to those of organic tissue) and because it allowed good visualization and measurement of crater characteristics. Depth of penetration, crater diameter, and extension of thermal damage were measured against power, focal length, and exposure time for each CO2 laser model. These results can be used as an index of behaviour of different surgical lasers. It appears that for fully characterizing the interaction of surgical lasers with the sample, it is necessary to specify either power, focal length, exposure time, or beam mode.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验