Hainsworth I R, McAllister E J, Simpson E, Duncan A, Roberts L B
J Clin Chem Clin Biochem. 1978 Apr;16(4):213-21. doi: 10.1515/cclm.1978.16.4.213.
The Jeol Clinalyzer was evaluated over a period of 15 weeks. The operating principles are briefly described. The functions of the mechanical components were tested and assessments made of the instrument's safety and reliability. The mechanical and electrical reliability of the instrument was excellent and the pumps and spectrophotometer gave good accuracy and precision. Between-batch precision of the analytical methods was good for urea, protein and bilirubin and acceptable for alkaline phosphatase and aspartate transaminase. There was a poor relative accuracy for alkaline phosphatase and aspartate transaminase and some proportional inaccuracy for urea and bilirubin.
对日本电子株式会社的临床分析仪进行了为期15周的评估。简要描述了其工作原理。测试了机械部件的功能,并对仪器的安全性和可靠性进行了评估。该仪器的机械和电气可靠性极佳,泵和分光光度计的准确性和精密度良好。分析方法的批间精密度对于尿素、蛋白质和胆红素而言良好,对于碱性磷酸酶和天冬氨酸转氨酶而言可接受。碱性磷酸酶和天冬氨酸转氨酶的相对准确性较差,尿素和胆红素存在一些比例误差。