终点离心分析法快速比浊法测定血清C3c和C4

Rapid turbidimetric determination of serum C3c and C4 by end point centrifugal analysis.

作者信息

Borque L, Preciado E, Zugaza C, Eugui J, Venta R

出版信息

Clin Biochem. 1983 Dec;16(6):330-3. doi: 10.1016/s0009-9120(83)80002-2.

Abstract

An end point turbidimetric method for the determination of C3c and C4 in serum using a centrifugal analyzer (Cobas Bio) is described. Several analytical factors were evaluated -pH, temperature, PEG and antibody concentration. Wide variations of temperature and pH did not significantly affect the turbidimetric reaction. A 20 g/L PEG concentration and 25-fold antiserum dilution were found satisfactory for the analysis. Precision of the assay was good and comparison with a RID method yielded an r value of 0.97. The procedure is simple and reliable.

摘要

描述了一种使用离心分析仪(Cobas Bio)通过终点比浊法测定血清中C3c和C4的方法。评估了几个分析因素——pH、温度、聚乙二醇(PEG)和抗体浓度。温度和pH的广泛变化对比浊反应没有显著影响。发现20 g/L的PEG浓度和25倍的抗血清稀释度用于分析是令人满意的。该测定方法的精密度良好,与放射免疫扩散法(RID)比较得出的r值为0.97。该方法简单可靠。