Dufrane S P, Malaisse W J, Sener A
Biochem Med. 1984 Aug;32(1):99-105. doi: 10.1016/0006-2944(84)90012-7.
A micromethod for the assay of aldose reductase is described. The method, which is based on the fluorometric measurement of the NADP+ formed when an aldose is converted to its corresponding polyol, was applied to lens and pancreatic islet crude homogenates, as well as semipurified lens aldose reductase. The fluorometric method has proved to be reproducible, more rapid, and more sensitive than the classical spectrophotometric procedure, and should find ready application in the screening of potential aldose reductase inhibitors.
本文描述了一种测定醛糖还原酶的微量方法。该方法基于对醛糖转化为相应多元醇时形成的NADP⁺进行荧光测定,已应用于晶状体和胰岛粗匀浆以及半纯化的晶状体醛糖还原酶。事实证明,该荧光法具有可重复性,比传统的分光光度法更快、更灵敏,应可在筛选潜在醛糖还原酶抑制剂方面得到广泛应用。