Dawson D J, Gribble M A
Med Phys. 1984 Sep-Oct;11(5):708-11. doi: 10.1118/1.595642.
A statistical analysis has been made of ionometrically measured parameters related to the central-axis output, the beam symmetry, the beam flatness, and the depth dose for the 6-MV x-ray beam from the Therac-6 linear accelerator. The means and coefficients of variation of the measurements are determined on a monthly basis for each parameter. Mean coefficients of variation less than +/- 1% have been determined and are useful in establishing the optimal limits of acceptance of these parameters. The results of this study indicate that the long-term stability of the radiation parameters associated with this unit can be maintained within acceptable limits with a consistent quality-assurance program.