Meiler J
Rontgenblatter. 1984 Jun;37(6):223-5.
It is shown that the Al equivalent of the prefiltration of an x-ray tube assembly varies significantly with the tube voltage, and hence the usual quotation of only one value without further specifications cannot suffice for the definition of the Al equivalent. The knowledge of the tube voltage, with which the equivalent was determined, is also important. The equivalent relationships can then be indicated for the total tube voltage range which is of interest.
结果表明,X射线管组件预过滤的铝当量随管电压显著变化,因此仅给出一个值而无进一步说明的通常报价不足以定义铝当量。确定当量时所使用的管电压信息也很重要。然后可以给出感兴趣的整个管电压范围内的当量关系。