Johnson C E, Cohen I A, Bickley S K, Hyder D M
Am J Hosp Pharm. 1984 Oct;41(10):2065-8.
The effect of various storage conditions on serum theophylline results using a fluorescence polarization immunoassay (FPI) method was studied. Two human subjects received sufficient oral theophylline to produce low and high therapeutic serum theophylline concentrations. Aliquots of venous whole blood (without anti-coagulant) and serum from each subject were sealed in polypropylene plastic and glass containers and stored at 4 degrees C and 25 degrees C. Additional aliquots of serum from each subject were stored in glass and polypropylene containers at -20 degrees C. FPI assays were performed on days 0 and 3 for room temperature serum samples, days 3 and 7 for refrigerated serum samples, and days 3, 7, 14, 28, 56, 168, and 336 for frozen serum samples. All whole blood samples were assayed on day 3 after removal of the clot. Some samples were tested using FPI for 1,3-dimethyluric acid, a metabolite of theophylline, and results from spiked samples and samples tested with high-pressure liquid chromatography were compared with FPI results to determine cross-reactivity of the metabolite. No appreciable change in the assayed theophylline values occurred under any storage temperature, nor was there any difference in results from samples stored in glass versus polypropylene containers. At a concentration of 20 micrograms/ml, 1,3-dimethyluric acid produced a cross-reactivity of 3.5% with the FPI system. The FPI method is not affected by storage of samples under the specified conditions over the time period studied.
研究了使用荧光偏振免疫测定法(FPI)时,各种储存条件对血清茶碱检测结果的影响。两名受试者口服足量茶碱,使血清茶碱浓度达到低治疗浓度和高治疗浓度。将每名受试者的静脉全血(无抗凝剂)和血清等分试样密封于聚丙烯塑料容器和玻璃容器中,分别储存在4℃和25℃。每名受试者的其他血清等分试样储存在玻璃容器和聚丙烯容器中,于-20℃保存。对室温血清样本在第0天和第3天进行FPI检测,对冷藏血清样本在第3天和第7天进行检测,对冷冻血清样本在第3天、第7天、第14天、第28天、第56天、第168天和第336天进行检测。所有全血样本在去除血凝块后第3天进行检测。一些样本使用FPI检测茶碱的代谢产物1,3 - 二甲基尿酸,并将加标样本和用高压液相色谱法检测的样本结果与FPI结果进行比较,以确定该代谢产物的交叉反应性。在任何储存温度下,检测的茶碱值均无明显变化,储存在玻璃容器和聚丙烯容器中的样本结果也无差异。在浓度为20微克/毫升时,1,3 - 二甲基尿酸与FPI系统的交叉反应性为3.5%。在所研究的时间段内,FPI方法不受指定条件下样本储存的影响。