可移除式高剂量铱-192脑植入物。犬脑的技术与体内测量

Removable high intensity iridium-192 brain implants. Technique and in vivo measurements in canine brain.

作者信息

Sewchand W, Amin P P, Drzymala R E, Salazar O M, Salcman M, Samaras G M, Botero E

出版信息

J Neurooncol. 1984;2(3):177-85. doi: 10.1007/BF00253269.

Abstract

A preclinical evaluation of the technical details and dosimetry for temporary high intensity 192Ir brain implants is presented. The canine brain was used for this quality assurance study in which direct in vivo dose measurements were done by thermoluminescent dosimetry (TLD rods). Precise and reproducible positioning of the TLD rods and 192Ir ribbons were assured by simple accessories which can be utilized in the clinical situation. The neurosurgical procedure for this non-routine interstitial implant of the brain, suitability of type and size of afterloading cannulas and facility for firmly anchoring them to the scalp, and comparison of measured doses with computer-predicted values are details assured by the canine study. Agreement between the in vivo determination and computer-generated doses was consistently in the range 2-5%. Data derived from this preclinical evaluation are currently used in both stereotactic and non-stereotactic brain implants at our institution. Details are presented for the implant procedure, dose measurements and brachytherapy planning for multiple ribbons. The latter incorporates direct interaction on computed tomography (CT) images for a hypothetical patient case.

摘要

本文介绍了临时性高强度192铱脑植入物的技术细节和剂量测定的临床前评估。本质量保证研究使用犬脑,通过热释光剂量测定法(TLD棒)进行直接体内剂量测量。简单的附件确保了TLD棒和192铱带的精确和可重复定位,这些附件可用于临床情况。犬类研究确保了这种非常规脑间质植入的神经外科手术、后装插管的类型和尺寸的适用性以及将其牢固锚定在头皮上的设施,以及测量剂量与计算机预测值的比较。体内测定与计算机生成剂量之间的一致性始终在2%-5%的范围内。本临床前评估得出的数据目前在我们机构的立体定向和非立体定向脑植入中均有应用。文中介绍了多根铱带的植入程序、剂量测量和近距离治疗计划的详细情况。后者纳入了针对一个假设患者病例在计算机断层扫描(CT)图像上的直接交互。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验