Parry B W, Anderson G A
Equine Vet J. 1984 Nov;16(6):529-31. doi: 10.1111/j.2042-3306.1984.tb02009.x.
Seventeen horses were used to determine the variances associated with blood pressure cuff application (Sp2) and with other inherent errors (So2). Systolic pressure values had Sp2 = 3.9 mmHg and So2 = 5.6 mmHg, while diastolic pressure values had Sp2 = 1.1 mmHg and So2 = 4.4 mmHg. Thus, to be considered different, two blood pressure means (in mmHg), each derived from three readings, had to differ by at least 3.9 for systolic pressure and 3.4 for diastolic pressure when all readings were made without cuff displacement; 6.8 for systolic pressure and 4.6 for diastolic pressure when the cuff was reapplied between, but not during, measurement of each mean; and 5.0 for systolic pressure and 3.8 for diastolic pressure when the cuff was reapplied between all readings. It was concluded that uniform cuff application is readily achieved.
使用17匹马确定与血压袖带应用相关的方差(Sp2)以及与其他固有误差相关的方差(So2)。收缩压值的Sp2 = 3.9 mmHg,So2 = 5.6 mmHg,而舒张压值的Sp2 = 1.1 mmHg,So2 = 4.4 mmHg。因此,要被认为有差异,当所有读数均在袖带无移位情况下进行时,两个血压均值(以mmHg为单位),每个均值由三次读数得出,收缩压至少相差3.9,舒张压至少相差3.4;当在测量每个均值之间(但不是在测量过程中)重新应用袖带时,收缩压相差6.8,舒张压相差4.6;当在所有读数之间重新应用袖带时,收缩压相差5.0,舒张压相差3.8。得出的结论是,很容易实现袖带的均匀应用。