[使用移动X射线设备的仰卧位胸部影像:软射线技术与硬射线技术的比较]

[Supine thoracic images with a mobile roentgen unit: comparison between the soft-ray and hard-ray technics].

作者信息

Prager P, Neumann D, Geiger K, Jaschke W, Diezler P, Brundin K

出版信息

Rontgenblatter. 1984 Dec;37(12):409-13.

PMID:6523005
Abstract

To compare the influence of kilovoltage on the quality of supine chest roentgenograms obtained with a mobile x-ray unit, we examined the films of 50 patients on whom supine chest radiographs had been taken incidentally within a period of 24 hours both with high and with low-kVp technique. For evaluation of the image quality, a total of 11 criteria were compared and quantitated by 5 observers using a simple score method. The same advantages that have been attributed to the high-kVp technique on upright films were found on the corresponding supine examinations. These films showed a more balanced penetration and greater sharpness than those taken with low-kVp. The wider latitude of the high-kVp technique simplifies the exposure, and fewer repeat takes will therefore be necessary. The diagnostic gain outweights the few disadvantages, which are mostly coupled to the increase in scattered radiation. This, as well as the availability of more powerful and efficient mobile generators, should promote routine application of high-kVp technique for supine chest examinations.

摘要

为比较千伏值对使用移动X线设备获得的仰卧位胸部X线片质量的影响,我们检查了50例患者的胸片,这些患者在24小时内分别用高千伏和低千伏技术偶然拍摄了仰卧位胸部X线片。为评估图像质量,5名观察者使用简单评分法对总共11项标准进行了比较和量化。在相应的仰卧位检查中,发现了与高千伏技术应用于立位片时相同的优势。这些片子比用低千伏拍摄的片子显示出更均匀的穿透性和更高的清晰度。高千伏技术更宽的宽容度简化了曝光过程,因此需要重复拍摄的次数更少。诊断上的获益超过了少数缺点,这些缺点大多与散射辐射增加有关。这一点,以及更强大、更高效的移动发生器的可用性,应会促进高千伏技术在仰卧位胸部检查中的常规应用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验