Boncompagni L, Del Chicca M G, Bonuccelli U, Clerico A
Boll Soc Ital Biol Sper. 1984 Sep 30;60(9):1743-9.
We used a new radioimmunological (RIA) kit for the assay of B subunit of creatine kinase enzyme (CK). This RIA system uses a specific antisera against the B subunit as ligant, human CK-BB labelled with 125I as tracer, and purified human CK-BB isoenzyme as standard. The mean (+/- SD) sensitivity obtained was 0.25 +/- 0.16 ng/tube and the between assay variability was about 9-10%. Serum levels of 113 normal subjects was not normally distributed. The 95% of values was found below 5 ng/ml. This new RIA is usefull in clinical practice when serum levels of CK-BB isoenzyme must be determined. This method is quickly and it is characterized by a good degree of precision, but the CK-MB isoenzyme cross-reacts for about 40% in this RIA system. Therefore, for the clinical diagnosis by means of this RIA it is necessary to rule out the concomitant elevations of serum CK-MB values.
我们使用了一种新的放射免疫分析(RIA)试剂盒来检测肌酸激酶(CK)的B亚基。该RIA系统使用针对B亚基的特异性抗血清作为配体,用125I标记的人CK-BB作为示踪剂,纯化的人CK-BB同工酶作为标准品。获得的平均(±标准差)灵敏度为0.25±0.16 ng/管,分析间变异性约为9-10%。113名正常受试者的血清水平呈非正态分布。发现95%的值低于5 ng/ml。当必须测定CK-BB同工酶的血清水平时,这种新的RIA在临床实践中很有用。该方法快速,具有良好的精密度,但在该RIA系统中,CK-MB同工酶的交叉反应约为40%。因此,通过这种RIA进行临床诊断时,有必要排除血清CK-MB值的同时升高。