Magn Reson Imaging. 1984;2(3):221-6. doi: 10.1016/0730-725x(84)90008-0.
The quality of nuclear magnetic resonance (NMR) image is degraded by a number of disturbing influences. In this article the influence of a nonlinear radiofrequency receiving amplifier is investigated. The effect of a cubic nonlinearity on the NMR signal is analyzed. Finally, the influence on the NMR image is studied for Fourier reconstruction techniques. As a result, the images are degraded by an additional term which is proportional to a parameter characterizing the nonlinearity. The additional term causes homogeneous objects to appear inhomogeneous and gives rise to image intensity outside the object. Under normal conditions the image degradation is negligible.
核磁共振(NMR)图像的质量会受到多种干扰因素的影响而下降。在本文中,研究了非线性射频接收放大器的影响。分析了三次非线性对NMR信号的作用。最后,针对傅里叶重建技术研究了其对NMR图像的影响。结果表明,图像会因一个与表征非线性的参数成比例的附加项而退化。该附加项会使均匀物体看起来不均匀,并导致物体外部出现图像强度。在正常情况下,图像退化可忽略不计。