Human gradient-layer calorimeter: development of an accurate and practical instrument for clinical studies.

作者信息

McManus C, Newhouse H, Seitz S, Nixon D, Poppendiek H, Heymsfield S

出版信息

JPEN J Parenter Enteral Nutr. 1984 May-Jun;8(3):317-20. doi: 10.1177/0148607184008003317.

Abstract

A whole body gradient-layer calorimeter was designed for thermophysiologic studies in hospitalized patients. The instrument was assembled in a room adjoining a metabolic ward. The patient lies recumbent on a stretcher which is wheeled into the calorimeter on tracks. Once inside, the patient's head extends through one wall into a clear plastic canopy. The canopy is partially open on one side and is the entry point for fresh air into this open-circuit system. Heat flux transducers in the calorimeter walls, a thermopile in the exit air stream, and a platemeter in conjunction with wet and dry bulb thermometers allow measurement of total, wet, and dry heat losses from the subject. All calorimeter output signals are routed to a microcomputer system which processes the data and provides real-time output of results. A mannequin heat source is used to calibrate the gradient layer. Testing with a variety of wet/dry heat sources indicate total, wet, and dry heat measurements are accurate to within 2-3%. A typical study of resting energy expenditure requires 1 hr, and protocols of up to 6 hr are well tolerated. The calorimeter has proven to be reliable, accurate, and easy to operate. These qualities, combined with good patient tolerance, make it ideal for metabolic studies.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验