Siggaard-Andersen O, Fogh-Andersen N, Thode J
Scand J Clin Lab Invest Suppl. 1983;165:43-6.
An effect of erythrocytes on the liquid junction potential of the junction "concentrated KCl whole blood" has previously been described, the effect being a positive bias of +1,4 mV with a high erythrocyte volume fraction (0.75). We have studied the erythrocyte effect with different concentrated salt solutions, and found a similar positive bias with KBr, KJ, KNO3, K2SO4, but a negative bias with NaCl, NaNO3, LiCl and CaCl2. The erythrocyte bias may be eliminated by mixing salts with opposite effect, e.g. KCl + NaCl (1.0 + 7.0 mol/l), KNO3 + NaNO3 (1.0 + 3.0 mol/l) KNO3 + Li-acetate (0.15 + 4.0 mol/l). The use of a mixed salt bridge solution may be advantageous for eliminating the erythrocyte bias when measuring ionic activity (e.g. Na+, K+, Ca2+) directly in whole blood.
此前已有关于红细胞对“浓氯化钾全血”液接电位影响的描述,在高红细胞体积分数(0.75)时,该影响为 +1.4 mV 的正偏差。我们用不同的浓盐溶液研究了红细胞效应,发现使用溴化钾、碘化钾、硝酸钾、硫酸钾时也有类似的正偏差,但使用氯化钠、硝酸钠、氯化锂和氯化钙时则为负偏差。通过混合具有相反效应的盐(例如氯化钾 + 氯化钠(1.0 + 7.0 mol/l)、硝酸钾 + 硝酸钠(1.0 + 3.0 mol/l)、硝酸钾 + 醋酸锂(0.15 + 4.0 mol/l)),可以消除红细胞偏差。在直接测量全血中的离子活性(例如 Na+、K+、Ca2+)时,使用混合盐桥溶液可能有利于消除红细胞偏差。