Matrai A, Flute P T, Dormandy J A
Biorheology Suppl. 1984;1:99-101. doi: 10.3233/bir-1984-23s116.
Aggregation and sedimentation of the red cells in the viscometer makes it difficult to obtain good reproducibility at low shear rates. A simple method is described to obtain repeatable torque-time curves with the Contraves LS30. Before starting rotation, the blood sample is agitated by an up and down movement of the bob. This results in identical (within 1-2% of the actual amplitude) repeated tracings even at the lowest shear rates, even with extremely time-dependent samples. The subsequent torque-time curve of the disaggregated and unsettled samples provides an experimental basis for the quantitative description of blood viscoelasticity and for dynamic analysis of red cell aggregation: furthermore it results in more accurate measurements when comparing different samples and testing therapeutic effects. The method is applicable to any coaxial cylinder type viscometer with manual lowering of the bob.
红细胞在粘度计中的聚集和沉降使得在低剪切速率下难以获得良好的重现性。本文描述了一种使用Contraves LS30获得可重复扭矩-时间曲线的简单方法。在开始旋转之前,通过摆动杆的上下移动来搅动血液样本。即使在最低剪切速率下,对于极其依赖时间的样本,这也能产生相同(在实际振幅的1%-2%范围内)的重复描记。随后,解聚且未沉降样本的扭矩-时间曲线为血液粘弹性的定量描述和红细胞聚集的动态分析提供了实验基础:此外,在比较不同样本和测试治疗效果时,它能带来更准确的测量结果。该方法适用于任何通过手动降低摆动杆的同轴圆筒型粘度计。