The halo vest: an evaluation of motion and forces across the neck.

作者信息

Koch R A, Nickel V L

出版信息

Spine (Phila Pa 1976). 1978 Jun;3(2):103-7.

PMID:663758
Abstract

Flexion-extension motion and compression-distraction forces across the cervical spine were evaluated in 6 patients in halo vests and in 1 patient in a halo cast. Motion in the vest was evaluated in supine and upright positions and averaged 31% of normal motion at the levels tested. Compression-distraction forces showed great individual variation and frequent compression. The average distraction force varied in different positions by nearly 20 pounds in the vest and over 30 pounds in the cast. Methods of improving cervical spine stability in halo orthoses are discussed.

摘要

对6例佩戴头环背心和1例佩戴头环石膏的患者进行了颈椎屈伸运动及压缩-牵张力评估。对头环背心的运动评估分别在仰卧位和直立位进行,测试节段的运动平均为正常运动的31%。压缩-牵张力表现出很大的个体差异且频繁出现压缩情况。牵张力平均值在不同体位下,头环背心相差近20磅,头环石膏相差超过30磅。文中讨论了改善头环矫形器中颈椎稳定性的方法。