Kellogg T F
Anal Biochem. 1983 Oct 1;134(1):137-43. doi: 10.1016/0003-2697(83)90275-0.
The effect of sample vial type and sample composition on the Cerenkov count rate detected from 32P and 36Cl was studied using a liquid scintillation counter. When counting was done in the noncoincident mode, glass vials allowed higher counting efficiency than plastic vials. In the coincident mode light scattering caused by polyethylene and polyproplyene vials allowed higher counting efficiency than glass vials. Highest coincident counting efficiency was from plastic minivials in a glass carrier vial. Increased solute concentration in samples caused increased counting efficiency due to changes in the refractive index of the solution. This can cause significant counting efficiency changes with no sample channel ratio change in density gradient fractions. The use of wavelength shifters is shown to be inappropriate when the sample pH varies, as this can change the fluorescent properties of the shifters and thereby the observed count rate.
使用液体闪烁计数器研究了样品瓶类型和样品组成对从³²P和³⁶Cl检测到的切伦科夫计数率的影响。在非符合模式下进行计数时,玻璃瓶比塑料瓶具有更高的计数效率。在符合模式下,由聚乙烯和聚丙烯瓶引起的光散射使得计数效率高于玻璃瓶。最高的符合计数效率来自玻璃载体瓶中的塑料微型瓶。样品中溶质浓度的增加由于溶液折射率的变化而导致计数效率提高。这可能导致在密度梯度级分中没有样品通道比变化的情况下计数效率发生显著变化。当样品pH值变化时,波长移位剂的使用被证明是不合适的,因为这会改变移位剂的荧光特性,从而改变观察到的计数率。