Ivan L P, Badejo A
Childs Brain. 1983;10(6):361-8. doi: 10.1159/000120137.
Using the Ladd fiberoptic system, fontanel pressure measurements were performed in infants from 1 to 219 days of age. Various holding devices were used for standardizing the application force and remarkably consistent values were found among three different holders. In group I (neonates), 9.45, 8.45 and 8.2 mm Hg were the mean values. In group II (4 days to 4 weeks), 10.3 and 9.2 were the mean values. In group III (34-219 days), the mean values were 9.9 and 12.4. In a group of 36 hydrocephalic children, epidural, ventricular and fontanel pressures were compared during surgery, and linear correlation was found but the ventricular fluid pressure readings were consistently lower. In dogs with artificial fontanels, cisternal and fontanel pressure was monitored while volume pressure studies were performed. In 80% of the cases, close linear correlation was found. Fontanel pressure measurement and monitoring give very valuable data on ICP. The correct zero reference point is a difficulty which can be eliminated with standardized holding devices.
使用拉德光纤系统,对1至219日龄的婴儿进行囟门压力测量。使用了各种固定装置来标准化施加力,并且在三种不同的固定装置中发现了非常一致的值。在第一组(新生儿)中,平均值分别为9.45、8.45和8.2毫米汞柱。在第二组(4天至4周)中,平均值为10.3和9.2。在第三组(34至219天)中,平均值为9.9和12.4。在一组36名脑积水儿童中,在手术期间比较了硬膜外、脑室和囟门压力,发现了线性相关性,但脑室液压力读数始终较低。在有人造囟门的狗中,在进行容量压力研究时监测了脑池和囟门压力。在80%的病例中,发现了密切的线性相关性。囟门压力测量和监测为颅内压提供了非常有价值的数据。正确的零参考点是一个难题,使用标准化固定装置可以消除这一难题。