过滤后测定羊水颗粒成分中“板层小体”磷脂的荧光测定法。

Fluorometric procedure for determining "lamellar body" phospholipids in the particulate fraction of amniotic fluid after filtration.

作者信息

Egberts J, Soederhuizen W, Gebhardt D O

出版信息

Clin Chem. 1983 Feb;29(2):362-5.

PMID:6687380
Abstract

A rapid (30-min) semiautomated continuous-flow procedure is described for use in assessing the phospholipids of the particulate ("lamellar body") fraction of human amniotic fluid. The method is based on measuring the difference in fluorescence of 1,6,-diphenyl-1,3,5-hexatriene added to amniotic fluid before and after micropore filtration. The filtration step removes "lamellar body" particles, which are considered to contain the fetal lung surfactant. The phospholipid values for the filtered particles are independent of background fluorescence, which increases when amniotic fluid is contaminated by bilirubin pigments or blood components. Over a wide range (3-150 mumols/L) the fluorescence increases linearly with the phospholipid concentration of the amniotic fluid. There is a good agreement between the value for particulate "lamellar body" phospholipid, the ratio of the "lamellar body" phospholipids to total amniotic fluid phospholipids, and the lecithin/sphingomyelin ratio.

摘要

本文描述了一种快速(30分钟)的半自动连续流动程序,用于评估人羊水颗粒(“板层小体”)部分的磷脂。该方法基于测量微孔过滤前后添加到羊水中的1,6 - 二苯基 - 1,3,5 - 己三烯荧光差异。过滤步骤去除了被认为含有胎儿肺表面活性物质的“板层小体”颗粒。过滤后颗粒的磷脂值与背景荧光无关,当羊水被胆红素色素或血液成分污染时,背景荧光会增加。在很宽的范围内(3 - 150μmol/L),荧光随羊水磷脂浓度线性增加。颗粒“板层小体”磷脂的值、“板层小体”磷脂与总羊水磷脂的比值以及卵磷脂/鞘磷脂比值之间具有良好的一致性。

相似文献

3
A comparison between the lecithin/sphingomyelin ratio of Gluck and other phospholipid values of amniotic fluid.
Am J Obstet Gynecol. 1975 Jul 15;122(6):734-6. doi: 10.1016/0002-9378(75)90579-7.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验