Cauberghs M, Van de Woestijne K P
Respiration. 1984;45(1):22-5. doi: 10.1159/000194592.
The forced oscillation technique developed by Korn et al. in 1979 (Siregnost FD 5) overestimates the measured resistances by about 20% in the range of resistance values met in healthy subjects and in patients. This is due to an overestimation of the impedance of the side tubing used for calibration. In addition, the device that we investigated introduces a phase shift of 9 degrees between pressure and flow.
1979年由科恩等人开发的强迫振荡技术(Siregnost FD 5)在健康受试者和患者所遇到的电阻值范围内,高估了测量电阻约20%。这是由于用于校准的侧管阻抗被高估。此外,我们所研究的设备在压力和流量之间引入了9度的相移。