Rehfeld S J, Barkeley J, Loken H F
Clin Chem. 1984 Feb;30(2):304-7.
A neutral ligand ion-exchanger type of ion-selective electrode is used in the current version of the Orion Model SS-20 calcium analyzer. We evaluated it over the pH range 4.0 to 8.5, using albumin solutions containing 150 mmol/L NaCl and pooled patients' serum. Within this pH range response was nernstian, and the relationship between potential (in millivolts) and log [Ca2+] at each pH was linear. Variation in Ca2+ with increasing ionic strength (because of added NaCl) was not significantly different between standards and albumin solutions; it decreased in both. In contrast, the proportion of calcium in serum increased with increasing NaCl concentrations. Ca2+ was not bound in whole serum albumin at a pH less than about 3.9. We observed a substantial difference between the degree of Ca2+ binding in solutions of fatty-acid-free human albumin and serum.
当前版本的奥立龙SS - 20型钙分析仪使用的是中性配体离子交换型离子选择性电极。我们在pH值4.0至8.5的范围内,使用含有150 mmol/L氯化钠的白蛋白溶液和混合的患者血清对其进行了评估。在此pH范围内,响应符合能斯特方程,并且在每个pH值下,电位(以毫伏为单位)与log [Ca2+]之间的关系呈线性。标准溶液和白蛋白溶液中,随着离子强度增加(由于添加了氯化钠),Ca2+的变化没有显著差异;两者均降低。相比之下,血清中钙的比例随着氯化钠浓度的增加而增加。在pH值小于约3.9时,Ca2+在全血清白蛋白中未结合。我们观察到在无脂肪酸的人白蛋白溶液和血清中,Ca2+的结合程度存在显著差异。