pH值和氯化钠对使用新型钙选择性电极测定血清和人白蛋白基质中离子钙的影响。

Effect of pH and NaCl on measurements of ionized calcium in matrices of serum and human albumin with a new calcium-selective electrode.

作者信息

Rehfeld S J, Barkeley J, Loken H F

出版信息

Clin Chem. 1984 Feb;30(2):304-7.

PMID:6692542
Abstract

A neutral ligand ion-exchanger type of ion-selective electrode is used in the current version of the Orion Model SS-20 calcium analyzer. We evaluated it over the pH range 4.0 to 8.5, using albumin solutions containing 150 mmol/L NaCl and pooled patients' serum. Within this pH range response was nernstian, and the relationship between potential (in millivolts) and log [Ca2+] at each pH was linear. Variation in Ca2+ with increasing ionic strength (because of added NaCl) was not significantly different between standards and albumin solutions; it decreased in both. In contrast, the proportion of calcium in serum increased with increasing NaCl concentrations. Ca2+ was not bound in whole serum albumin at a pH less than about 3.9. We observed a substantial difference between the degree of Ca2+ binding in solutions of fatty-acid-free human albumin and serum.

摘要

当前版本的奥立龙SS - 20型钙分析仪使用的是中性配体离子交换型离子选择性电极。我们在pH值4.0至8.5的范围内,使用含有150 mmol/L氯化钠的白蛋白溶液和混合的患者血清对其进行了评估。在此pH范围内,响应符合能斯特方程,并且在每个pH值下,电位(以毫伏为单位)与log [Ca2+]之间的关系呈线性。标准溶液和白蛋白溶液中,随着离子强度增加(由于添加了氯化钠),Ca2+的变化没有显著差异;两者均降低。相比之下,血清中钙的比例随着氯化钠浓度的增加而增加。在pH值小于约3.9时,Ca2+在全血清白蛋白中未结合。我们观察到在无脂肪酸的人白蛋白溶液和血清中,Ca2+的结合程度存在显著差异。