Suppr超能文献

一款离子化镁-酸碱度分析仪——NOVA 8的评估

Evaluation of an ionized magnesium-pH analyzer--NOVA 8.

作者信息

Thode J, Juul-Jørgensen B, Seibaek M, Elming H, Borresen E, Jordal R

机构信息

Department Clinical Biochemistry, Gentofte Hospital, University of Copenhagen, Hellerup, Denmark.

出版信息

Scand J Clin Lab Invest. 1998 Apr;58(2):127-33. doi: 10.1080/00365519850186706.

Abstract

An ion-selective analyzer (NOVA 8 [previously NOVA CRT], NOVA Biomedical Waltham, USA) for simultaneous measurement of the concentration of ionized magnesium (cMG2+) and pH is investigated for linearity and influence of calcium ions on the Mg2+ results in different Mg2+ aqueous solutions with/without added calcium. Within the cMg2+ range 0-3.0 mmol/L, which covers most clinical values, we found cMg2+ values close to the line of identity, with a sensitivity about 94%, but 0.09 mmol/L lower in the absence of Ca2+. The analytical precision (CV) for human serum control was 1.7% in the physiological range. The accuracy and precision data for the Mg2+ electrode meet the demands for reliable results within the range for clinical use. We found a reference interval (95%) for cMg2+ of between 0.43 and 0.59 mmol/L serum, with a mean value of 0.51 mmol/L for healthy adults in the non-fasting state with no relation to sex or age. The mean cMg2+ was 61% of the mean concentration of plasma total magnesium (cTMg) ranging from (95%) 49% to 73%. No significant correlation was found between cMg2+ and total protein, albumin, phosphate, Ca2+, and total CO2. No significant difference was found between mean values for cMg2+ in whole blood, plasma, and serum. Serum samples could be stored for 24 h at 4 degrees C without significant change. Storage at -20 degrees C for 7 months caused a mean cMg2+ decrease of 8.3%. The mean slope delta lgcMg2+/delta pH measured after equilibration with different pCO2 values in each serum sample from the reference population was -0.0110, indicating a close agreement between the values for actual cMg2+ and adjusted cMg2+ (pH = 7.4).

摘要

研究了一种用于同时测量离子化镁(cMG2+)浓度和pH值的离子选择性分析仪(NOVA 8[以前的NOVA CRT],美国马萨诸塞州沃尔瑟姆市的NOVA Biomedical公司),考察其在添加/未添加钙的不同镁离子水溶液中的线性以及钙离子对镁离子结果的影响。在0-3.0 mmol/L的cMg2+范围内(涵盖了大多数临床值),我们发现cMg2+值接近恒等线,灵敏度约为94%,但在无Ca2+时低0.09 mmol/L。人血清对照的分析精密度(CV)在生理范围内为1.7%。镁离子电极的准确度和精密度数据满足临床使用范围内可靠结果的要求。我们发现血清中cMg2+的参考区间(95%)为0.43至0.59 mmol/L,非空腹状态下健康成年人的平均值为0.51 mmol/L,与性别或年龄无关。cMg2+的平均值为血浆总镁(cTMg)平均浓度的61%,范围为(95%)49%至73%。未发现cMg2+与总蛋白、白蛋白、磷酸盐、Ca2+和总CO2之间存在显著相关性。全血、血浆和血清中cMg2+的平均值之间未发现显著差异。血清样本在4℃下可保存24小时而无显著变化。在-20℃下保存7个月导致cMg2+平均下降8.3%。在参考人群的每个血清样本中,用不同pCO2值平衡后测得的平均斜率δlgcMg2+/δpH为-0.0110,表明实际cMg2+值与调整后的cMg2+值(pH = 7.4)之间高度一致。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验