Allen W C, Heiple K G, Burstein A H
J Bone Joint Surg Am. 1978 Jun;60(4):506-15.
An intramedullary rod for the femur was designed and compared with rods commonly used. Its bending strength was increased 40 to 80 per cent and its torsional rigidity was increased 230 to 3,000 per cent compared with the other devices tested. In forty patients with traumatic and pathological fractures and osteotomies of the femur, excellent fixation was achieved and very early weight-bearing was possible. In all fresh fractures and non-infected non-unions, union occurred very promptly and bone-grafting was not required in most instances. The stregth and rigidity of the device made it exceptionally useful for non-unions, pathological fractures of the femur, and resection-fusions of the knee.