脊髓监测作为脊髓诱发电位的临床应用。

Spinal cord monitoring as a clinical utilization of the spinal evoked potential.

作者信息

Tamaki T, Noguchi T, Takano H, Tsuji H, Nakagawa T, Imai K, Inoue S

出版信息

Clin Orthop Relat Res. 1984 Apr(184):58-64.

PMID:6705365
Abstract

For the purpose of monitoring spinal cord function, evoked spinal cord potential was observed by recording electrodes in the subarachnoid space at the level of the conus medullaris and stimuli delivered to the rostral spinal cord from the epidural space. A specially designed, flexible tube-type electrode was used for both recording and stimulating. The spinal cord evoked potential consisted of an initial spike followed by polyphasic waves. The peak amplitude change was considered to be the main indicator of events that would potentially be hazardous to spinal cord function. Two hundred and twenty-nine operations were monitored by the recording electrode. Six of the patients experienced postoperative neurologic complications. During operation, the peak amplitudes of the evoked potentials diminished to less than 50% of the control values. Simultaneous monitoring of evoked potential revealed that this method was more sensitive than the wake-up test. No complications resulted from the application of this method. Thus, monitoring is a sufficiently sensitive, safe, and reliable method to be useful during aggressive surgery in the spine and spinal cord.

摘要

为监测脊髓功能,通过在脊髓圆锥水平的蛛网膜下腔放置记录电极,并从硬膜外腔向脊髓头端施加刺激来观察脊髓诱发电位。使用一种特别设计的柔性管型电极进行记录和刺激。脊髓诱发电位由一个初始尖峰和随后的多相波组成。峰值幅度变化被认为是可能对脊髓功能有潜在危害事件的主要指标。用记录电极监测了229例手术。其中6例患者术后出现神经并发症。手术过程中,诱发电位的峰值幅度降至对照值的50%以下。诱发电位的同步监测表明,该方法比唤醒试验更敏感。应用该方法未产生并发症。因此,监测是一种足够敏感、安全且可靠的方法,在脊柱和脊髓的积极手术中很有用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验