Early results of valve replacement with the Björk-Shiley convexoconcave prosthesis.

作者信息

Marshall W G, Kouchoukos N T, Pollock S B, Bradley E L

出版信息

Ann Thorac Surg. 1984 May;37(5):398-403. doi: 10.1016/s0003-4975(10)60764-3.

Abstract

The Björk-Shiley convexoconcave prosthetic valve has design characteristics that may result in a lower incidence of thromboembolic complications than the conventional spherical Björk-Shiley prosthesis. We evaluated the results of valve replacement with the convexoconcave prosthesis in 248 patients receiving 301 prosthetic valves between March, 1979, and June, 1981. One hundred thirteen patients had aortic valve replacement (AVR), 73 had mitral valve replacement (MVR), and 62 had multiple valve replacement. Two hundred nine (84%) were in New York Heart Association Class III or IV. The median duration of follow-up was 13 months, and follow-up information was available for 246 (99%) of the patients. The actuarial incidence of freedom from thromboembolism at two years was 98% in the AVR group, 97% in the MVR group, and 87% in the group having multiple valve replacement. There were no documented episodes of valve thrombosis or mechanical failure and no fatal thromboembolic complications. The absence of valve thrombosis is in marked contrast to the results reported with the spherical disc valve. Although longer follow-up is necessary, it appears that the convexoconcave design represents a major improvement in the Björk-Shiley prosthesis.