Yoganathan A P, Strand D M, Woo Y R, Stevenson D M
Med Instrum. 1984 Mar-Apr;18(2):109-13.
An on-line digital method is described for analyzing the pressure drop and regurgitative characteristics of prosthetic heart valves in a pulse duplicator system. The method is based on the Apple II microcomputer system but could be modified to be used with most microcomputers currently available. The digital data collection system is synchronized with the pulse duplicator system, and is programmed to collect the relevant pressure (upstream and downstream) and volumetric flow data for 10 cardiac cycles at a given time. The system is relatively easy to use and gives the scientist a fair amount of flexibility in terms of data collection, storage, and analysis. The on-line method has been used with the pulse duplicator of the Georgia Institute of Technology for the past 3 years.
描述了一种在线数字方法,用于分析脉动复制器系统中人工心脏瓣膜的压降和反流特性。该方法基于苹果II微机系统,但可进行修改以与目前可用的大多数微机配合使用。数字数据采集系统与脉动复制器系统同步,并编程为在给定时间采集10个心动周期的相关压力(上游和下游)和容积流量数据。该系统相对易于使用,并且在数据采集、存储和分析方面为科学家提供了相当大的灵活性。在过去3年中,该在线方法已与佐治亚理工学院的脉动复制器一起使用。