Kemp G J
Clin Chem. 1984 Jul;30(7):1168-70.
I describe three ways of using data from analytical-recovery studies to assess analytical nonlinearity, without access to samples of known concentration. I then describe a recovery-based method of assessing constant, proportional, and non-linear errors with use of as little as one sample pool of known concentration. In each case, I present the theoretical basis of the method and an outline of a practical experimental protocol.