Regional cerebral blood flow measurements using stable xenon enhanced computed tomography: a theoretical and experimental evaluation.

作者信息

Kishore P R, Rao G U, Fernandez R E, Keenan R L, Arora G D, Gadisseux P, Stewart L M, Wist A O, Fatouros P P, Dillard D

出版信息

J Comput Assist Tomogr. 1984 Aug;8(4):619-30. doi: 10.1097/00004728-198408000-00007.

Abstract

Several theoretical and practical aspects of regional cerebral blood flow measurements using stable xenon gas and CT are discussed. It is shown that by comparing the enhancement at any time T1 with that at saturation or any other time T2, the need to use arbitrary means to bring the arterial concentration data and the CT enhancement data to the same system of measurement units can be eliminated. If CT is performed continuously during the washin phase, say at intervals of 1 min, least squares analysis of the enhancement data can be used to obtain the best possible estimates for the flow rate constant kappa and the saturation enhancement. However, if only a limited number of scans can be performed, as may be the case in human studies, it is also possible to get a good estimate of kappa from a knowledge of the ratio of the enhancement at any time T1 with that at any other time T2. Combinations of T1 = 2.0 min and T2 = 4.0 min, T1 = 1.0 min and T2 = 6.0 min, or T1 = 2.0 min and T2 = 5.0 min were found to be the most convenient. It is also shown that the end-tidal xenon concentration in the exhaled air can be accurately assessed indirectly by measuring the oxygen, CO2, and water vapor concentrations, thereby eliminating the need for more expensive methods involving the use of a mass spectrometer or a thermal conductivity gas analyzer.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验