Suppr超能文献

[99mTc-pertechnetate neck uptake in the evaluation of thyroid function--simple method with gamma camera and pinhole collimator (author's transl)].

作者信息

Maeda T, Kawakami M, Seto H, Aburano T, Yoshizaki T, Hisada K

出版信息

Radioisotopes. 1978 May;27(5):241-4. doi: 10.3769/radioisotopes.27.5_241.

Abstract

The neck uptake ratio of 99mTcO4- was measured using a gamma camera and pinhole collimator. The patient was sit 2 m apart from the collimator surface and radioactivity was counted 20 minutes after injection. The whole body count with lead shield (Cc) and without shield (Cd), and natural background count (CBG) were measured. As the shield was utilized the B filter for conventional thyroidal 131I uptake test (lead plate of 1.3 cmx10 cmx10 cm) and it was placed over the anterior neck to cover the thyroid gland at the measurement of (Cc). Then 99mTcO4- neck uptake ratio was calculated as following: (formula: see text) The mean values were 5.3%, 4.5%, 3.1%, 11.2% and 20.7% in the group of euthyroid, hypothyroid, subacute thyroiditis, hyperthyroid state with chronic thyroiditis and Graves' disease, respectively. This method is very convenient and simple to perform following scintigraphy of the thyroid gland with pinhole collimator.