Lewy A J, Markey S P
Science. 1978 Aug 25;201(4357):741-3. doi: 10.1126/science.675255.
Three techniques have been used to measure human plasma melatonin: bioassay, radioimmunoassay, and gas chromatography--mass spectrometry (GC-MS). GC-MS is theoretically capable of the greatest specificity, but in general suffers from insufficient sensitivity. Negative chemical ionization, a new technique, provides a 150-fold increase in GC-MS sensitivity for electron-capturing compounds. Negative chemical ionization GC-MS permits routine measurement in human plasma of melatonin at a concentration as low as 1 picogram per milliliter.
生物测定法、放射免疫测定法和气相色谱 - 质谱联用(GC-MS)法。理论上,GC-MS法具有最高的特异性,但一般灵敏度不足。负化学电离是一项新技术,可使GC-MS对电子捕获化合物的灵敏度提高150倍。负化学电离GC-MS能够对人体血浆中的褪黑素进行常规测量,其浓度低至每毫升1皮克。