Boyer A L, Goitein M
Med Phys. 1980 Jan-Feb;7(1):19-26. doi: 10.1118/1.594654.
A technique has been developed and implemented by which individualized Ellis compensating filters can be rapidly and accurately produced from Moiré topographic photographs. A novel Moiré camera has been designed such that it has been possible to mount it on a radiotherapy simulator. The camera does not interfere with normal use of the simulator but can be deployed as needed. Depth of missing tissue contours are produced by the camera over the treatment field, accurately registered to the patient's simulated treatment position. An optical enlargement technique is used to transfer the depth of missing tissue contours in the photograph to a lead sheet compensating filter.
一种技术已被开发并实施,通过该技术可以从莫尔条纹地形图快速准确地制作个体化的埃利斯补偿滤过器。设计了一种新型莫尔条纹相机,使其能够安装在放射治疗模拟机上。该相机不会干扰模拟机的正常使用,但可根据需要进行部署。相机在治疗野上生成缺失组织轮廓的深度,并精确对准患者的模拟治疗位置。采用光学放大技术将照片中缺失组织轮廓的深度转移到铅片补偿滤过器上。