Lakier J B, Khaja F, Magilligan D J, Goldstein S
Circulation. 1980 Aug;62(2):313-8. doi: 10.1161/01.cir.62.2.313.
An analysis of 211 patients who had porcine xenograft valve replacements at Henry Ford Hospital between October 1971 and March 1974, was accomplished, with 100% follow-up. The follow-up period extended from 60-89 months after implantation. One hundred sixty-seven patients with 192 valves survived the perioperative period and were subjected to life table analysis. Hemodynamically significant porcine xenograft degeneration that required reoperation occurred in 18 patients, two of whom had infective endocarditis. Only four valves failed within 48 months of surgery. Ten of 42 (23.8%) patients with isolated aortic valve replacement and eight of 102 patients (7.8%) patients with isolated mitral valve replacement required reoperation (p < 0.01). In patients under 25 years of age, six of nine surviving patients had repeat operations. Our data indicate that porcine xenograft degeneration is related to the duration of implantation and the age of the patient at the time implantation was performed. In addition, porcine xenograft valves in the aortic position are more likely to degenerate than are those in the mitral position.
对1971年10月至1974年3月期间在亨利·福特医院接受猪异种移植瓣膜置换术的211例患者进行了分析,随访率达100%。随访期为植入后60至89个月。167例植入192个瓣膜的患者度过围手术期并接受寿命表分析。18例患者发生了需要再次手术的具有血流动力学意义的猪异种移植瓣膜退变,其中2例患有感染性心内膜炎。仅4个瓣膜在术后48个月内出现故障。42例(23.8%)孤立主动脉瓣置换患者中有10例,102例孤立二尖瓣置换患者中有8例(7.8%)需要再次手术(p<0.01)。在25岁以下的患者中,9例存活患者中有6例接受了再次手术。我们的数据表明,猪异种移植瓣膜退变与植入持续时间及植入时患者年龄有关。此外,主动脉位置的猪异种移植瓣膜比二尖瓣位置的更易退变。