Kirsch C M, Moore S C, Zimmerman R E, English R J, Holman B L
J Nucl Med. 1981 Aug;22(8):726-31.
We evaluated a single-photon emission computed tomographic system using ten scanning detectors in a circular array. The system uses focusing collimators that scan radially as well as tangentially. The spatial resolution in the tomographic plane is 2.6 cm FWHM and the axial resolution is 3.3 cm FWHM. The resolution is independent of position within the field of view, and nearly independent of energy through 511 keV. Sensitivity was found to be 4600 cps/muCi-ml for an extended phantom, 20-cm in diameter, filled with Tc-99m; 7200 cps/muCi-ml with TI-201; and 8000 cps/muCi-ml with Ga-67. Investigations of positional uniformity indicated some quantitative distortion of data due to inadequate attenuation correction. Improvement in the attenuation correction is necessary before truly quantitative tracer distribution studies are undertaken.
我们评估了一个采用十个扫描探测器呈圆形阵列排列的单光子发射计算机断层扫描系统。该系统使用聚焦准直器,可进行径向和切向扫描。断层平面内的空间分辨率为半高宽2.6厘米,轴向分辨率为半高宽3.3厘米。分辨率与视野内的位置无关,并且在511千电子伏特能量范围内几乎与能量无关。对于一个直径20厘米、填充有Tc-99m的扩展模型,灵敏度为4600计数每秒/微居里-毫升;对于TI-201为7200计数每秒/微居里-毫升;对于Ga-67为8000计数每秒/微居里-毫升。位置均匀性研究表明,由于衰减校正不足,数据存在一些定量失真。在进行真正的定量示踪剂分布研究之前,必须改进衰减校正。