J Biochem Biophys Methods. 1982 Jun;6(2):173-7. doi: 10.1016/0165-022x(82)90063-x.
A new design of the observation window to observation chamber seal is described for the Berger rapid stopped-flow apparatus. This design permits reliable operation of the apparatus and retains the features of a square non-deviating observation path with flush windows. Incorporation of fiber optic light guides for the observation path has made the apparatus insensitive to vibration artifacts. A demonstration reaction of calcium binding to EGTA is presented; the reaction, which is 90% complete within 2 ms, illustrates the capability of this apparatus for examining fast reactions.