Malfunction in a St. Jude medical mitral valve prosthesis: a clinical diagnostic dilemma.

作者信息

Kafka H, Parker J O, Charrette E J, Salerno T A

出版信息

Can J Surg. 1983 Jan;26(1):84-6.

PMID:6821772
Abstract

Thrombosis of one leaflet of a St. Jude medical prosthetic mitral valve in a 49-year-old man resulted in prosthetic malfunction causing cardiac failure. Clinical and echocardiographic assessment did not adequately distinguish between abnormal valve function and left ventricular dysfunction as the responsible lesion. Hemodynamic study revealed a gradient across the mitral valve and poor left ventricular function. At operation it was found that thrombus had caused one of the leaflets to stick in an almost open position. This malfunction was attributed to lack of adequate anticoagulation and to the impingement of a piece of papillary muscle on the prosthesis. A definite echocardiographic pattern for malfunction of the St. Jude valve has not been established. However, it is clear that loudness of prosthetic valve sounds cannot be used as a guide to normal function in a bileaflet valve.

摘要

一名49岁男性的圣犹达医疗公司人工二尖瓣的一个瓣叶发生血栓形成,导致人工瓣膜功能障碍,进而引发心力衰竭。临床和超声心动图评估未能充分区分异常瓣膜功能和作为病因的左心室功能障碍。血流动力学研究显示二尖瓣跨瓣压差以及左心室功能不佳。手术时发现血栓已导致其中一个瓣叶卡在几乎开放的位置。这种功能障碍归因于抗凝不足以及一块乳头肌对人工瓣膜的撞击。圣犹达瓣膜功能障碍的明确超声心动图模式尚未确立。然而,很明显,双叶瓣膜中人工瓣膜声音的响度不能用作正常功能的判断依据。