Preoperative wire localization of mammographically identified nonpalpable breast lesions is facilitated by using a perforated mammographic compression plate. This device is used in conjunction with a self-retaining wire, and its use results in accurate placement of the wire-marker tip in close proximity to the mammographically suspicious lesion. In 39 of 47 lesions, the wire tip was within 1.0 cm of the lesion, and in 45 of 47 lesions, the wire tip was within 1.5 cm of the lesion.
使用穿孔乳腺摄影压迫板有助于对乳腺摄影发现的不可触及乳腺病变进行术前导丝定位。该装置与自固定导丝配合使用,其使用可使导丝标记尖端准确放置在乳腺摄影可疑病变附近。在47个病变中的39个中,导丝尖端位于病变1.0厘米范围内,在47个病变中的45个中,导丝尖端位于病变1.5厘米范围内。