Early experience with a universal (DDD) pacing device.

作者信息

Hayes D L, Holmes D R, Vlietstra R E

出版信息

Mayo Clin Proc. 1983 May;58(5):301-6.

PMID:6843180
Abstract

To assess the advantages and complications of a new universal (DDD) pacemaker, we studied retrospectively the initial 38 patients who received such a pacing device. This group consisted of 27 men and 11 women whose ages ranged from 23 to 89 years. The pacemaker was the initial one for 32 patients and was the replacement unit for 6. Indications for dual-chamber pacing included a need to maintain atrioventricular synchrony, the previous occurrence of the pacemaker syndrome or the presence of intact ventriculoatrial conduction, and previous success with atrioventricular sequential pacing. Pacemaker-mediated tachycardia developed in six patients during the follow-up period. All six were successfully treated by a change in pacing mode. Eighteen pacemakers remained in the DDD mode, 17 were reprogrammed to the DVI mode, and 3 were reprogrammed to the VVI mode. Seven patients experienced difficulties with loss of capture or undersensing, the most common problem being failure of atrial sensing. This was corrected in four of five patients by noninvasive reprogramming. Pacemaker-mediated tachycardia is a frequent complication that can be avoided by assessing ventriculoatrial conduction before implantation of the pacemaker. Use of the DDD device should be considered in patients in whom atrioventricular synchrony must be maintained or who have previously had the pacemaker syndrome with VVI pacing.