Suppr超能文献

CYBEST 2型模型用于实际妇科肿块筛查的实验进展报告。标本拒收阈值和标本制备的改变。

Progress report on experimental use of CYBEST model 2 for practical gynecologic mass screening. Alterations of the specimen rejection threshold and specimen preparation.

作者信息

Mukawa A, Kamitsuma Y, Jisaki F, Tanaka N, Ikeda H, Ueno T, Tsunekawa S

出版信息

Anal Quant Cytol. 1983 Mar;5(1):31-4.

PMID:6846967
Abstract

In order to maintain an acceptable performance of CYBEST model 2 in the field of gynecologic mass screening, the algorithm of rejection for the CYBEST assessment of inadequately oligocellular samples was reviewed, and the mean number of detected cells per square millimeter for passing rejection was raised from two cells to four. The results of field tests obtained after the modification showed a false-positive rate of 28.3% for 9,010 samples as compared with 41.8% for 10,549 samples before the modification while the rejection rate showed a moderate yet acceptable increase after the modification from 4.7% to 7.0%. Some modifications of the specimen preparation routine were also introduced to improve the performance of the system.

摘要

为了在妇科肿块筛查领域保持CYBEST 2型模型的可接受性能,对CYBEST评估细胞数量不足样本时的排除算法进行了审查,通过排除的每平方毫米检测细胞平均数从两个细胞提高到了四个。修改后进行的现场测试结果显示,9010个样本的假阳性率为28.3%,而修改前10549个样本的假阳性率为41.8%,同时排除率在修改后有适度但可接受的提高,从4.7%提高到了7.0%。还对样本制备程序进行了一些修改,以提高系统性能。